- Description
-
Details
Battery Powered High Frequency Generators vs Single or 3 Phase HF Generators
Our battery powered generators range from 32kW to 50kW and plug right into a standard outlet. In case electrical conversion to power an x-ray generator proves too costly, this might be a cost-effective alternative. Not sure how long you will be leasing your current office space and don't want to spend twice in electrical work? Ask our experts about this solution.
All our x-ray configurations can be used with the battery operated generator; from your stand-up or traditional x-ray to your integrated floor fail system or straight arm
- Additional Information
-
Additional Information
Specifications Constant Potention Power Ratings 32 kW 40 kW 50 kW Radiographic kVp Range in 1 kVp Steps Accuracy 40 - 125/150
±(3% + 1 kV)mA Range & Stations 10, 12.5, 16, 20, 25, 32, 40,
50, 64, 80, 100, 125, 160,
200, 250, 320, 400
±(4% + 1 mA)10, 12.5, 16, 20, 25, 32, 40,
50, 64, 80, 100, 125, 160,
200, 250, 320, 400, 500
±(4% + 1 mA)10, 12.5, 16, 20, 25, 32, 40,
50, 64, 80, 100, 125, 160,
200, 250, 320, 400, 500
±(4% + 1 mA)Power Output 400 mA @ 80 kVp
320 mA @ 100 kVp
250 mA @ 125 kVp
200 mA @ 150 kVp500 mA @ 80 kVp
400 mA @ 100 kVp
320 mA @ 125 kVp
250 mA @ 150 kVp500 mA @ 100 kVp
400 mA @ 120 kVp
320 mA @ 138 kVp
250 mA @ 150 kVpExposure Time Range Accuracy 0.001-10 seconds
±(2%+0.1ms)mAs Range 0.1 - 500 mAs (*)s High Voltage Ripple (TYP) <1 kVp @ 100 kVp Automatic Exposure Control (up to 4 chambers) Optional Tomo Optional Buckys (2 units standard) Optional up to 4 Anatomical Programming Optional Uniterruptible Power Supply (1) Optional Fluoroscopic kVp Range in 1 kVp Steps Accuracy 40 - 120
±(3%+1 kV)Fluoroscopy
Fixed PPS
Variable PPS(optional)at 50Hz at 60 Hz
50 PPS 60 PPS
1, 2, 3, 6, 1, 2, 4, 8,
12, 25, 50 15, 30, 50Automatic Brightness Control Optional 3/2 Point Operation Console 3 P.Standard / 2 P.Optional Line Voltage Range and Phases Single Phase Line Automatic Regulation from 100 - 240 VAC (±10%) Storage Capacity 67,500 mAs @ 100 kVp Minimum Power Requirments According to IEC 601-2-7 = 2,200 W (X-Ray Tube + Generator)
(1) Uninterruptible Power Supply Option - Line power requirement = 500 W (X-Ray Tube + Generator
(*)640 mAs - on request - REQUEST QUOTE
-